Carry out box flushing before sensor removal when the sensor has been in process for
several months. Box flushing removes dried process medium from isolation valve and makes
sensor removal easier.
WARNING Do
not activate box steam flush, if the sensor and the retractor tool are not installed to the
isolation valve.
Close the ¼ " valve to the nozzle (1).
Open the ¼ " valve to the box (2).
In the DTR transmitter, go to MENU > SENSOR STATUS and activate wash by pressing WASH button. Repeat
the wash 3 … 5 times.