| WARNING Always shut the main steam
valve before performing any work on the wash nozzle. |
Check the nozzle and the valve before installing the wash nozzle. Use thread seal tape for
all thread connections.
-
Insert the nozzle into the isolation valve (1).
-
Attach the nozzle to the nozzle guide with one special M5x10 screw (2) using a 5 mm
Allen key.
-
Remove the safety pin (3).
-
Close the ¼" check port valve under the nozzle isolation valve (1).
-
Open the isolation valve (2) by turning the handle counterclockwise.
-
Push the nozzle to the process (3).
-
Attach the nozzle to the nozzle guide with one s pecial M5x10 screw (4) using a
5 mm Allen key.
-
Lock the isolation valve handle with the safety pin (5).
-
Connect the lines in the following order.
-
STEAM: Connect the steam line and sensor flush flexible line to the nozzle T-piece
(1,2).
-
WATER: Connect the water line to ¼" valve (1,2).
-
Open the steam supply line valve (3).
-
Check the functionality of washing from the optical image window in DTR.