PR‑23‑RP specifications - PR-23 - PR-23-AC - PR-23-AP - PR-23-GC - PR-23-GP - PR-23-M - PR-23-RP - PR-23-SD - PR-23-W

PR-23 Series User Guide

Document code
IM-EN-PR23
Revision
K
Language
English
Product
PR-23
PR-23-AC
PR-23-AP
PR-23-GC
PR-23-GP
PR-23-M
PR-23-RP
PR-23-SD
PR-23-W
Document type
User guide
General specifications

Feature

Specification

Refractive index range

Full range nD 1.3200 … 1.5300, sapphire prism

Accuracy

Refractive index nD ±0.0002 (corresponds typically to ±0.1 % by weight)

Repeatability and stability correspond to accuracy

Speed of response

1 s undamped, damping time selectable up to 5 min

Calibration

With Cargille certified refractive index liquids over full range of nD 1.3200 … 1.5300

CORE‑Optics

No mechanical adjustments

Digital measurement

3648 pixel CCD element

Light source

LED 589 nm wavelength, sodium light

Temperature sensor

Built‑in Pt‑1000

Temperature compensation

Automatic, digital compensation

Instrument verification

With certified refractive index liquids and Vaisala documented procedure

Ambient temperature

−20 °C … +45 °C (−4 °F … +113 °F)

Sensor PR‑23‑RP specifications

Feature

Specification

Sensor PR‑23‑RP Probe type, seamless one-piece wetted parts construction with no welds

Process connection

Flanges: ANSI 2 in 300 lbs

Process pressure

Up to 25 bar (350 psi)

Process temperature

−40 °C … +150 °C (−40 °F … +302 °F)

Process wetted parts, standard

AISI316L Stainless steel, Alloy 20 or Alloy C276; prism sapphire, prism gaskets PTFE

Sensor protection class

IP67, Nema 4X

Protection class Protection class 3

Sensor weight

7.89 kg (17.4 lbs)