PR‑23‑GC mounting specifications - PR-23 - PR-23-AC - PR-23-AP - PR-23-GC - PR-23-GP - PR-23-M - PR-23-RP - PR-23-SD - PR-23-W

PR-23 Series User Guide

Document code
IM-EN-PR23
Revision
K
Language
English
Product
PR-23
PR-23-AC
PR-23-AP
PR-23-GC
PR-23-GP
PR-23-M
PR-23-RP
PR-23-SD
PR-23-W
Document type
User guide

The compact refractometer is mounted either in a pipe elbow by an L coupling connection or in a straight pipe using a wafer flow cell or a pipe flow cell. Both flow cell mounting designs create optimum flow velocity on the measurement surface providing a good self-cleaning effect. The wafer flow cell has an optional automatic wash system capability.

In a pipe of at least 2.5 in diameter the sensor is mounted in a pipe elbow with an L coupling. In a 2 in pipe the sensor is mounted in a pipe elbow using reducing ferrule PR-9283. In 0.5 in, 1 in and 1.5 in pipes a flow cell is installed in a straight pipe. The flangeless wafer flow cell is a compact alternative to traditional flow cells. The wafer refers to a flow cell body that is installed between DIN, ANSI or JIS piping flanges with bolts and nuts. The wafer flow cell is a one‑piece body construction with no welds. A pipe flow cell is also available for 0.5 in and 1 in pipes.

Mounting sensor in pipe 2.5 in or larger
Mounting sensor in 2 in pipe
Mounting sensor with PFC flow cell
Mounting sensor with WFC flow cell
Mounting wafer flow cell and sensor in vertical pipe
Mounting wafer flow cell and sensor in horizontal pipe