Feature |
Specification |
---|---|
Refractive index range |
Full range nD 1.3200 … 1.5300 (corresponds to hot water – 100 Brix) |
Accuracy |
Refractive index nD ±0.0002 (corresponds typically to ±0.1 % by weight) Repeatability and stability correspond to accuracy |
Speed of response |
1 s undamped, damping time selectable up to 5 min |
Calibration |
With Cargille certified refractive index liquids over full range of nD 1.3200 … 1.5300 |
CORE‑Optics |
No mechanical adjustments |
Digital measurement |
3648 pixel CCD element |
Light source |
LED 589 nm wavelength, sodium light |
Temperature sensor |
Built‑in Pt‑1000 |
Temperature compensation |
Automatic, digital compensation |
Instrument verification |
With certified refractive index liquids and Vaisala documented procedure |
Ambient temperature |
−20 °C … +45 °C (−4 °F … +113 °F) |
Feature |
Specification |
---|---|
Sensor PR‑23‑AC |
Compact sensor model for small pipelines |
Process connection |
Sanitary clamp 2.5 in; Varinline® in‑line access unit clamp DN65 or using elbow flow cell (for line sizes of 2.5 in and smaller) |
Sanitary design |
3‑A Sanitary Standard 46‑04 approved and EHEDG (European Hygienic Engineering & Design Group) certified (connections ‑H, ‑E) |
Process pressure |
Sanitary clamp max. 15 bar (200 psi) at +20 °C (+70 °F) / 9 bar (125 psi) at +120 °C (+250 °F) |
Process temperature |
−20 °C … +130 °C (−4 °F … +266 °F) |
Process wetted parts, standard |
AISI 316 L stainless steel, spinel prism, prism gaskets MTF (Modified Teflon) |
Sensor protection class |
IP67, Nema 4X |
Protection class | Protection class 3 |
Sensor weight |
2.0 kg (4.4 lbs) |