Feature |
Specification |
|---|---|
| Refractive index range, standard |
Full range nD 1.3200 … 1.5300, sapphire prism |
| Refractive index range, option | nD 1.2600 … 1.4700, sapphire prism H74 |
|
Accuracy |
Refractive index nD ±0.0002 (corresponds typically to ±0.1 % by weight) Repeatability and stability correspond to accuracy |
|
Speed of response |
1 s undamped, damping time selectable up to 5 min |
|
Calibration |
With Cargille certified refractive index liquids over full range of nD 1.3200 … 1.5300 |
|
CORE‑Optics |
No mechanical adjustments |
|
Digital measurement |
3648 pixel CCD element |
|
Light source |
LED 589 nm wavelength, sodium light |
|
Temperature sensor |
Built‑in Pt‑1000 |
|
Temperature compensation |
Automatic, digital compensation |
|
Instrument verification |
With certified refractive index liquids and Vaisala documented procedure |
|
Ambient temperature |
−20 °C … +45 °C (−4 °F … +113 °F) |
|
Feature |
Specification |
|---|---|
| Sensor PR‑23‑MS | Teflon® body sensor model for aggressive medium |
|
Sensor protection class |
IP67, Nema 4X |
| Protection class | Protection class 3 |
|
Sensor weight |
5.0 kg (12.1 lbs) |
|
Feature |
Specification |
|---|---|
|
Process connection |
Thread G ½ in or ½ in NPT female |
|
Process pressure |
Max. 10 bar (145 psi) |
|
Process temperature |
Max. +130 °C (+266 °F) |
|
Process wetted parts, standard |
Teflon® (PTFE) or Kynar® (PVDF), prism gaskets MTF (Modified Teflon®), sapphire prism, O‑ring Kalrez, adaptor sapphire |