Verification procedure - PR-23 - PR-23-AC - PR-23-AP - PR-23-GC - PR-23-GP - PR-23-M - PR-23-RP - PR-23-SD - PR-23-W

PR-23 Series User Guide

Document code
IM-EN-PR23
Revision
K
Language
English
Product
PR-23
PR-23-AC
PR-23-AP
PR-23-GC
PR-23-GP
PR-23-M
PR-23-RP
PR-23-SD
PR-23-W
Document type
User guide
  1. To start the verification process, select Main > VERIFICATION.

    The first verification display instructs you on the pre-verification procedure:

  2. When you are finished with the preparations, press CONTINUE (rightmost soft key) to start the verification process.
  3. The verification itself is done by the refractometer system. Follow the instructions on screen and apply one R.I. liquid at a time on the sensor and press VERIFY (rightmost soft key).
    Clean and dry the prism and the sample holder very carefully between the R.I. liquids. Use a suitable solvent, for example ethanol or IPA (isopropyl alcohol).
  4. To check that the standard liquid is properly wetting the prism, press soft key OPT. IMAGE.
    The optical image should show a sharp shadow edge, see the following figure.
  5. A verification data collection method is implemented in the DTR. The instrument measures each verification data point 10 times and uses the average of these measurements. Measuring each verification liquid takes approximately 10 s, during which the measurement progress display is shown. Wait until the VERIFICATION STEP 2 display reappears before proceeding to the next verification liquid.

    The sample holder keeps the sample on the prism surface and also blocks the ambient light from reaching the prism.

  6. Press COMPLETE in the verification display to finish the verification procedure and to call up the verification results.

If verification is successful, meaning that all measurements are within ±0.0004 of the nominal values, message VERIFICATION OK is shown.

Verification completed successfully (here only with one R.I. liquid)
The sensor verification concerns only the refractive index nD measurement. The calculation of concentration from nD and process temperature TEMP is not included. For more information, see Calibrating concentration measurement.

If the verification fails, see Troubleshooting sensor verification.