Diagnostic values - PR-23 - PR-23-AC - PR-23-AP - PR-23-GC - PR-23-GP - PR-23-M - PR-23-RP - PR-23-SD - PR-23-W

PR-23 Series User Guide

Document code
IM-EN-PR23
Revision
K
Language
English
Product
PR-23
PR-23-AC
PR-23-AP
PR-23-GC
PR-23-GP
PR-23-M
PR-23-RP
PR-23-SD
PR-23-W
Document type
User guide

The values at the left of the graph are used for diagnostic purposes:

  • CONC is the final concentration value including field calibration adjustment, see Concentration calibration layers.
  • TEMP, see Temperature measurement.
  • CCD gives the position of the shadow edge on CCD in %.
  • nD is the refractive index value nD from the sensor.
  • CALC is the calculated concentration value without field calibration adjustment, Field calibration.
  • QF or Quality Factor is a value in the range of 0 to 200. The QF value depends on process medium optical properties. A typical good value is 100, but there are process media where 50 is acceptable. The QF value should stay constant during the process. Falling QF value may indicate prism coating.
  • LED is a measure of the amount of light from the light source in %.
  • HD TMP is the sensor head temperature, see Temperature measurement.
  • HD HUM is the sensor head humidity, see Sensor head humidity.
  • I_SNS value shows the current to sensor, the nominal value is 40 mA.
  • DTR TMP is the indicating transmitter temperature, see Temperature measurement.
  • DTR V1 gives the voltage from the power module, the nominal value is 24 V.
  • DTR V2 gives the DC supply voltage, the nominal value is 3.3 V.
The slope display also has a soft key SENSOR RESTART which can be used to restart the current sensor (see upper left corner of the display for sensor letter) after a sensor software update.