Sensor wiring must follow drawing WRG‑367 or WRG‑350, see Safe sensor wiring. See also Wiring for a prism wash system for steam and Wiring for prism wash system for high pressure water.

WARNING The FM unit installations must comply with the relevant requirements of National Electrical Code (ANSI/NFPA 70) for Division 2 Hazardous (Classified) Locations and all instructions in this manual. All wiring of PR‑23‑…‑FM systems must run in a conduit.
Safe sensor wiring
CAUTION Do not modify the unit or use it in ways not described in the documentation. Improper modification or use may lead to safety hazards, equipment damage, failure to perform according to specification, or decreased equipment lifetime.
CAUTION Do not connect or disconnect the sensor connector when the circuits are energized. Switch off the power from indicating transmitter DTR external power switch before disconnecting the sensor cable from the sensor. After connecting the sensor cable back to the sensor, you can switch power back on.
CAUTION Make sure that the rated voltage is not exceeded by transient disturbances of more than 119 V.
WARNING The refractometer models PR‑23‑M, PR‑23‑MS and PR‑23‑W contain parts made of PTFE in sensor heads. These parts are subject to electrostatic hazard. The refractometer models PR‑23‑M, PR‑23‑MS and PR‑23‑W must only be used for measuring liquids with high conductivity (> 10000 pS/m) in the presence of hazardous atmospheres.
WARNING The painted surface of the enclosure of PR‑23‑MS refractometer is subject to electrostatic hazard. Clean only with a damp cloth in the presence of hazardous atmospheres.