Technical Data - HMM100

HMM100 Quick Guide

Document code
M211002EN
Revision
E
Language
English
Product
HMM100
Document type
User guide
Property Description/Value
Measured Parameters and Measurement Range

Relative humidity

Dew point

0 … 100 %RH

−20 … +100 °C (−4 … +212 °F) Td

Accuracy (Including Non-Linearity, Hysteresis, and Repeatability)

at −20 … +40 °C (−4 … +104 °F)

±2 %RH (0 … 90 %RH)

±3 %RH (90 … 100 %RH)

at −40 … −20 °C, +40 … +180 °C (−40 … −4 °F, +104 … +356 °F)

±2.5 %RH (0 … 90 %RH)

±3.5 %RH (90 … 100 %RH)

Humidity Sensor Types

HUMICAP® 180R

HUMICAP® 180

Analog Output Types (1 Output Selectable)

2-wire model

3-wire model

4 … 20 mA (loop-powered)

0 … 20 mA, 0 … 1/5/10 V, 1 … 5 V

Supply Voltage

2-wire model

3-wire model (0 … 20 mA, 0 … 1 V/5 V, and 1 … 5 V)

3-wire model (0 … 10 V)

24 VDC

10 … 35 VDC or 24 VAC ±20 %

15 … 35 VDC or 24 VAC ±20 %

Input Current 6 mA
Connections

Screw terminals

Service cable connector

0.5 … 1.5 mm2

M8 4-pin male

Probe Types and Cable Lengths

Stainless steel probe

Plastic probe (standard body)

Plastic probe (extended 400‑mm (16‑in) body)

0.6/1.55/2.9 m (2/5.1/9.5 ft)

0.6/1.55/2.9 m (2/5.1/9.5 ft)

1.55 m (5.1 ft)

Operating Temperature

Component board

Stainless steel probe and plastic probe (standard body)

Plastic probe (extended 400‑mm (16‑in) body)

Plastic grid filter and membrane filter

Porous PTFE filter, stainless steel sintered filter, and PPS plastic grid with stainless steel net filter

−5 … +55 °C (+23 … +131 °F)

−70 … +180 °C (−94 … +356 °F)

−70 … +120 °C (−94 … +248 °F)

−20 … +80 °C (−4 … +176 °F)

−70 … +180 °C (−94 … +356 °F)

Figure 1. HMM100 Dimensions. Top to Bottom: Standard and Extended Probe Body
Download the complete HMM100 User Guide from www.vaisala.com/hmm100