The Errata provides additional information about OPC UA 1.2 product functionality.
| Description | Workaround | Fix Version |
|---|---|---|
| Connecting to the OPC-UA server doesn't work with either OPC-generated or self-signed certificates or key pairs. | Use a username and password key pair to connect to the OPC-UA server. | |
| Some measurement units are not updated on the OPC-UA server after they are changed in viewLinc. | Restart the OPC-UA Server. | |
| Some Location properties are not updated on the OPC-UA client after they are changed in viewLinc. | Restart the OPC-UA Server. | |
| The OPC-UA unexpectedly restarts when connecting to viewLinc 5.1 Service Update 3 and 4. | Upgrade to a newer version of viewLinc. | viewLinc 5.1 Service Update 5. |