RFL100 firmware version 1.7.0 - RFL100

RFL100 Firmware Release Notes

Document code
DOC246158
Revision
N
Language
English
Product
RFL100
Document type
Release notes

Maintenance release.

Release date: May 2023

Fixes

  • Fixed a software task timing issue that affected the reliability of the wireless link between the data logger and the AP10 access point. The issue sometimes caused the data logger to disconnect from the access point and other rare instabilities.
  • Fixed an issue that caused the local display on the RFL100 to round measurement readings differently from the readings shown in viewLinc. For example, the RFL100 display may have shown a reading of 21.3 °C, while viewLinc, configured with one decimal resolution, showed 21.4 °C. With this fix, the RFL100 display will now show the same rounded measurement readings as viewLinc, providing consistent and accurate readings across the monitoring system.

Compatibility

Compatibility with AP10 firmware version 4.5.0 is added. Otherwise, same as version 1.6.0.