Support for Taiwan regional variant.
Release date: November 2021
New features and improvements
- Support for an NCC-certified Taiwan radio variant with identifier of 922 MHz.
Fixes
- Fixed a rare issue where the RFL100 could transmit some samples with wrong values into viewLinc.
- Data loggers without probes are now prevented from joining viewLinc (first time joining). There must be at least one probe attached to the logger to join the viewLinc Enterprise Server.
- Fixed an issue where upgrading the firmware in dual T mode could result in temperature measurement not being available on channel 2.
Compatibility and upgradeability
Compatibility with AP10 firmware version 4.1.0 is added. Otherwise, same as version 1.4.0.