DSC74 sampling cell with quick connector and leak screw - DMT132 - DMT143 - DMP8 - DPT146 - DMP80 - DMP7 - DMT152

Dew Point Sampling Cells for Vaisala DRYCAP® Products Technical Note

Product status
Document code
B212382EN-C
ft:locale
en-US
Product
DMT132
DMT143
DMP8
DPT146
DMP80
DMP7
DMT152
Document type
Technical note
DSC74 sampling cell with accessories
1
Quick connector
2
Leak screw
3
Sampling cell body (DMT242SC)
4
Connection point for probe/transmitter, ISO G1/2"
5
Thread adapter, type G3/8" - G1/4"
6
Thread adapter, type R3/8" - R1/2"

The DSC74 has been designed especially for compressed air lines. The sampling cell contains an adjustable leak screw that maintains process pressure while allowing airflow past the sensor for a representative measurement.

The leak screw is opened and closed with a screwdriver. The leak screw should be 1/2 turn open. This can be verified by first closing the leak screw, then turning 1/2 turn to open it. The air coming out will make a barely audible hiss and can be lightly felt when putting a hand in front of the air stream.

Do not open the leak screw more than 1/2 turn to avoid reducing the pressure in the sampling cell.

The DSC74 comes with a quick connector that fits to industry standard compressed air line connectors (suitable for type D, Quick08, NIP08). This allows for easy installation and detachment of the dew point transmitter without having to shut down the process. Alternative ways to connect are through the two different thread adapters (R3/8" to R1/2" and G3/8" to G1/4" ISO) that are supplied with each DSC74 unit.

When using the DSC74, seal the threads of the quick connector or thread adapter carefully with PTFE thread seal tape. Tighten with an open-end wrench.